Point to Point Insulation
Intrinsic Error or Reference Conditions or Code Max Error
Influence Quantity Rated Operating Conditions Seen
Intrinsic Error Reference Condions A 5%
Position Reference Positions ±90° E
1
N/A
Supply Voltage At the Limits Specified E
2
1.6%
by the Manufacturer
Temperature 0°C and 35°C E
3
1%
Absorbed Current of Maximum Absorbed Current
Device on Test of the Device on Test E
4
N/A
Low-frequency Low-frequency
Magnetic Field Magnetic field 0.4 kA/m E
5
N/A
Not Used Not Used E
6
N/A
Capacitance Cmax = 1uF E
7
2%
Waveform R.M.S. CR <= 2 E
8
N/A
Operating Error B=\A\+1.15
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
7.17%
PrimeTest 200
Operating Instructions
20
Differential Leakage
Intrinsic Error or Reference Conditions or Code Max Error
Influence Quantity Rated Operating Conditions Seen
Intrinsic Error Reference Condions A 5%
Position Reference Positions ±90° E
1
N/A
Supply Voltage At the Limits Specified E
2
0.56%
by the Manufacturer
Temperature 0°C and 35°C E
3
0.81%
Absorbed Current of Maximum Absorbed Current
Device on Test of the Device on Test E
4
1.5%
Low-frequency Low-frequency
Magnetic Field Magnetic field 0.4 kA/m E
5
N/A
Not Used Not Used E
6
N/A
Capacitance Cmax = 1uF E
7
N/A
Waveform R.M.S. CR <= 2 E
8
2.5%
Operating Error B=\A\+1.15
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
8.54%
Touch Leakage
Intrinsic Error or Reference Conditions or Code Max Error
Influence Quantity Rated Operating Conditions Seen
Intrinsic Error Reference Condions A 10%
Position Reference Positions ±90° E
1
N/A
Supply Voltage At the Limits Specified E
2
2.7%
by the Manufacturer
Temperature 0°C and 35°C E
3
1.04%
Absorbed Current of Maximum Absorbed Current
Device on Test of the Device on Test E
4
0.87%
Low-frequency Low-frequency
Magnetic Field Magnetic field 0.4 kA/m E
5
N/A
Not Used Not Used E
6
N/A
Capacitance Cmax = 1uF E
7
N/A
Waveform R.M.S. CR <= 2 E
8
0.79%
Operating Error B=\A\+1.15
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
13.59%