PrimeTest 250
+
Operating Instructions
25
Note: The power switch on the appliance
under test must be in the ON position
to perform an insulation test. If no
appliance is detected the PrimeTest
250
+
will display the following.
3.3 Testing a Class II Appliance
• Visually inspect the appliance and mains cord
for signs of damage.
• If the appliance passes a visual inspection
proceed with the electrical tests.
• Plug the earth test lead into the Earth Continuity
Test Terminal on the PrimeTest 250
+
end panel.
• Plug the appliance into the PrimeTest 250
+
front panel test/mains outlet socket.
• Connect the earth test probe to an exposed
metal part on the appliance.
• If the Appliance under test has an ON/OFF
switch, make sure it is in the ON position.
• Press the CLII test key.
• The display will momentarily show the insulation
test voltage before proceeding with the test.